Skip to content

SIMSEF Workflow

Workflow description

The spatial ion mobility-scheduled exhaustive fragmentation (SIMSEF) module can be used in trapped ion mobility (TIMS) imaging workflows to schedule fragmentation spectra acquisitions for every detected feature across the tissue. This achieves high MS2 coverage and allows more confident compound annotations. To use the SIMSEF tool, the prmMALDI prototype instrument control software is required, which is distributed by Bruker Daltonics to cooperation partners. You can contact your Bruker representative to ask for the prmMALDI prototype and refer to SIMSEF.

Requirements

  • access to timsTOF fleX instrument
  • access to
  • the prototypic timsControl 4.1/5.0 with the prm-MALDI option (contact your Bruker representative)
  • timsControl 5.1 with prmMALDI enabled (contact your Bruker representative)
  • Download SIMSEF acquisition tool for the respective timsControl version

Procedure

Note: ⚠ acquire MS1 data with a smaller laser spot size than raster size (= stage movement/pixel size) Note: ⚠ Due to the size of IMS-MS imaging data, we recommend to use 10,000-40,000 pixels in your MS1 run for SIMSEF experiments

  • this leaves some tissue for MS2 acquisition

  • Prepare the sample following your default procedure.

  • Setup the imaging run in flexImaging. Set the raster size bigger than the laser spot size in your imaging method (e.g. laser spot size = 20 µm, raster size = 50 µm).
  • Save the imaging run via "Save imaging run as...". This will store the geometry files for later use.
  • Restart timsControl (prm-MALDI prototype) to make the instrument recognize the geometry files. Recalibrate the instrument (recommended).
  • Start the MS1 imaging run.
  • Copy the raw data to a powerful data analysis machine (Windows and Linux support Bruker raw data)
  • Analyse the MS1 data and filter it to your liking ( see imaging worflow). If your IMS seperation allows, filter for isotopic peaks.
  • Run the SIMSEF scheduler. Using the preview, evaluate the fragmentation schedules and optimise the parameters. Check the task manager for running tasks. The scheduling may need some time.
  • Copy the schedule to the instrument computer. (Must be the same path as selected on the analysis computer, not including spaces or special characters)
  • Load the instrument method, you want to acquire the MS2 spectra with. Make sure the method is appropriate for your expected fragment ion m/z range.
    • Disable MALDI mode, enable PASEF mode, ensure that "advanced" is disabled for the collision energy settings.
    • Enable MALDI mode, save the method.
    • Set MS mode to MS/MS in the MALDI settings.
    • Increase the laser size to the raster size
  • Set instrument to operate and recalibrate if needed (recommended).
  • Select the correct geometry file on the instrument computer (saved in step 3)
  • Run "simsef_pewpew.exe" on the instrument computer, select the acquisition.txt and click "Run acquisition."
  • After the acquisition, copy the folder to the analysis computer.
  • Import all MS2 files
  • Run Mass detection with the centroid mass detector and noise level 50-100 (starting point, for MS2 only!)
  • Pair the MS2 spectra to the image features using the Assign MALDI MS2s to features module
  • Annotate using any annotation module in MZmine

Steffen Heuckeroth


Last update: July 31, 2024 14:02:36